This web site is designed for people who are setting up a Magnetic Resonance Imaging (MRI) scanner for functional imaging (fMRI). The basic idea of fMRI is to rapidly acquire T2*-weighted images of the brain while the participant performs a cognitive task. We can then analyze the data to determine which brain regions are changing their signal in response to the behavioral task. To do this, we typically need a way to present task (usually a visual task via a computer screen or an auditory task via headphones) while we collect behavioral data (e.g. button presses). We also need a way to synchronize the scans with the behavioral task.

Since MRI is an inherently expensive tool, most new centers should certainly budget for professional MRI-compatible solutions for stimuli presentation, behavioral recording and scan synchronization. I try to describe several of the professional options on this web page. However, it should be noted that the costs of MRI is largely fixed, and many scientists hope to utilize a clinical system when it is not required for clinical duties. For this audience, I describe inexpensive methods to solve these problems. However, I wish to stress that the implementation of these techniques must be done carefully, and I take no responsibility for any issues that might arise using the methods described here.

Aspects of MRI make it inherently unsuitable for many off-the-shelf electronic devices. Issues include a strong constant (static) magnetic field, rapid switching of magnetic fields (gradients, that can induce electrical signals), radio frequency (RF) transmission and RF acquisition.

Setup

I really urge new fMRI users to get to know their MRI service engineers and application specialists. These individuals are very knowledgable, and in my experience they are pivotal to getting the best out of your scanner. Further, it is worth giving them a heads up if you are going to add new hardware. Specifically, they can help ensure you have the correct software licenses (most users will need a Echo Planar Imaging [EPI] license, which is generally common and pre-installed). Second, you will want to have an idea of where the EPI trigger hardware is located – for Siemens scanners you will want an optical cable routed from the equipment room to the location where your behavioral computer is located (last I checked, the Philips MRI systems have an electrical connector at the console). You will also want the service engineer to help you enable scanner to generate the EPI trigger signals (on Siemens scanners, this is done once in the service software, whereas for Philips scanners this is done on the protocol scan card).

If you are establishing a new MRI center, it is critical to establish the location and configuration of your penetration panel(s). A penetration panel allows cables and light to enter in the scanner bay without introducing RF artifacts. The image on the right shows a typical research fMRI penetration panel with a computer projector, optical cables (PST response glove cables are black, scanner trigger pulse is orange), and some metal electrical cables (plugged into BNC jacks). Tubes (wave guides) that run through the penetration panel need to be tuned for the frequencies used by your scanner. The larger the bore of the wave guide, the longer (and heavier) the guide. It is easy to send light (e.g. a computer projector beam) or optical cables through a wave guide, but you need to be very careful putting any wire cables through a wave guide: the cables need to be shielded, the signals should not be at the frequency used by the scanner, the shield should be grounded, and typically the signals should be run through an RF filter (the photo shows some filters on the DB9 and DB25 jacks). It is generally a good idea to have wave guides professionally installed – if they are poorly mounted they can introduce RF artifacts into all your scans. For most clinical setups, there are usually a few unused waveguides you can exploit, but you will want to have a careful discussion with your service engineer to ensure that these wave guides are not intended for some future hardware.

Panel

You will want to carefully select the sequences used for your fMRI study. Generally, it is a good idea to carefully consult with a MRI physicist, and see examine the sequences used by other centers that have similar hardware. Here are examples of our teams evaluation of Diffusion-weighted imaging (DWI) and Arterial Spin Labeling (ASL) .

Visual stimuli

Computer screens are the typical way for presenting fMRI tasks. This allows you to present experiments written in EPrime, Presentation, Cogent, or your favorite programming language. There are basically four options here:

projector_location

The first three methods require a mirror mounted on the head coil. Your scanner manufacturer should provide you with one of these (seen on the headcoil in the picture on the right). The first two methods require a back projection screen. The image on the right shows the professional screen we use. However, a low budget version can be made using plastic (PVC) pipe from you local plumbing store and a large sheet of drafting mylar.

Auditory stimuli

projector_sm

Tactile stimuli

You can present tactile stimuli by using air puffs, piezoelectric ceramics, or electrical currents. However, each of these techniques has its own challenges and are used less often in everyday research. Therefore, I do not discuss these in detail here.

Eye Tracking

The human eye has poor acuity away from fixation, so you can learn a lot by observing where someone is looking. There are no simple, inexpensive solutions, but there are numerous professional solutions that you can find with a web search for the terms “fmri eye tracker”. In my experience, there is a huge variability in the professional products available. Before purchasing a system I strongly advise visiting a research site that has the intended system installed and seeing it in operation by scientists (rather than sales reps).

Recording button responses

Recording physiological data

pulse_0

Scanner Synchronization

trigger-300x150.jpg

One critical requirement is knowing the timing between behavioral events (when did the participant see/hear/do something) and fMRI acquisition. We typically want our behavioral experiment to start precisely when the scanner begins acquiring data. This turns out to be somewhat tricky, as the scanner initializes most scans with a shimming sequence (that can take a variable time) and also discards the first few scans (as these have more T1-effects). Therefore, starting your scanner when you hear the auditory sounds of the scanner or when you detect RF signals is not sufficient. Fortunately, modern scanners can be set to generate trigger signals at the beginning of each EPI volume (as described in the Setup section). The trick is turning these very brief (nano-second) signals into longer signals that most computer inputs can detect (milliseconds). Further, Siemens scanners generate optical signals, which we need to convert to electrical signals.

Several professional behavioral input devices include a trigger detector. For example, the Current Designs hardware includes an optical detector, while the PST Fiber Optic Button Response system includes an electrical trigger detector (via a BNC connector).

nano

You can also build your own trigger pulse detector. The photograph and schematics on the right show a very simple device using the ubiquitous TLC555 timer as a monostable multivibrator and the versatile U-HID nano . The photo shows the standard Siemens compatible optical connector on the left, this board contains a small light that is red when the power is on, but briefly flashes off when a scan pulse is detected. The circuit is socketed onto a UHID nano (right side) that supplies power and transmits responses to the computer via the USB cable. This circuit is easy to build on a bread board or solder onto a prolect board, but if you are interested you can also contact me for Gerber files that you can have fabricated by vendors such as the terrific OSH Park (USA) or Fritzing (Europe, they also provide excellent software for designing your own boards). I like the UHID devices because you can program them to emulate many different devices (mouse buttons, keyboard presses, gamepad presses) and once this configuration is programmed any computer will see it as the designated response. Whereas the professional devices have a fixed mapping (e.g. for PST, a trigger pulse emulates the ‘^’ keypress), the UHID can be any device. Personally, I prefer to emulate devices such as gamepads that do not fill up the keyboard buffer or interfere with any programs that may be running. More details regarding synchronizing with your scanner can be found at my data logging page.